Validation plays a crucial role in ensuring the quality of computer system performance. It enhances system reliability, reducing errors and minimizing risks to both processes and data integrity. Additionally, computer system validation helps lower long-term system and project costs by reducing maintenance and rework expenses.
The validation of an HPLC (High-Performance Liquid Chromatography) system holds particular significance due to its scrutiny by regulatory authorities, including the FDA, where data integrity issues have been identified during inspections. Therefore, HPLC serves as an illustrative example for system validation. The process of creating a computer system validation protocol can be broken down into the following steps:
- User Requirement Specification (URS): Prepare a URS for the HPLC system, outlining critical functionalities required for analysis. Consider the following points:
- Method creation and updating capabilities (including injection volume, flow rate, runtime, column temperature, wavelength, etc.).
- Sequence creation and updating capabilities (including injector requirements, vial selection, sequence copy-paste, etc.).
- Ability to run full or partial injection sequences.
- Report printing with customizable layouts and printer command options.
- Support for single injections without sequence preparation.
- User access control.
- Audit trail and electronic signature features for 21CFR compliance.
- Other requirements as necessary for the company or quality control department.
- Design Qualification: Ensure that the purchased computer system aligns with the URS and is suitable for connecting to the associated instrument. Inspect computer hardware, including CPU, connections, power supply, and switches.
- Verify the operating system configuration meets minimum requirements.
- Record instrument details such as name, model, serial number, software version, and instrument ID.
- Installation Qualification: During installation qualification, follow these steps:
- Compare the computer system and components with the purchase order.
- Examine all documents, including manuals, safety and validation certificates, and maintenance instructions.
- Inspect the system for any damage incurred during transit.
- Install the system by connecting its components.
- Power on the system and ensure all components pass electronic self-tests.
- Install the software following the manufacturer&39;s instructions.
- Verify the correct software installation.
- Configure printers and equipment modules.
- Document all hardware and software installed on the computer along with descriptions.
- Prepare an installation report, detailing hardware and software serial and revision numbers, hard disk and RAM sizes, operating software versions, monitor and printer serial numbers, manufacturer information, and network connections.
- Operational Qualification: Perform operational qualification for both the computer system and software:
- Validate computer boot-up, date and time settings, and hard disk functionality.
- Check printer drives, network connections, software versions, and program alterations.
For software qualification:
- Verify user authentication, access control, password policies, and audit trail functionality.
- Assess the number of password characters, failed login detection, user lockout, password expiry alarms, and system policy compliance.
- Test the audit trail feature.
For application software functionality, evaluate method preparation, method editing/deletion, operation qualification, and post-analysis processes (including integration and report generation).
Customize checkpoints based on software and instrument functionality, maintaining essential validation checkpoints specific to the software&39;s purpose.
- Backup and Restore: Take data backups after printing reports and chromatograms. After backing up, delete data files from the computer system, including the recycle bin. Restore deleted data from the backup file and compare the printed reports and chromatograms with the originals.
- Validation Report: Compile a comprehensive computer system validation report containing all observations, data, screenshots, and prints. Document any identified deficiencies and corresponding corrective actions.
- Training: Provide training to personnel responsible for operating the computer system and software. Maintain records of training attendance.
- Conclusion: Create a final conclusion report confirming adherence to all computer system validation steps and verifying the satisfactory functioning of all software features.
In summary, computer system validation is a rigorous process that ensures the reliable performance of critical systems, such as HPLC. By following these structured steps, organizations can maintain compliance with regulatory standards and ensure the accuracy and integrity of their data and processes.